YUSHI UM-1 Series Ultrasonic Thickness Gauge

Add time:2025-02-08 17:21

  • File size: 382.5KB

UM-1 Series Technical Parameters

 Display Screen  128×64 Dot Matrix LCD Display (EL Backlight)
 Working Principle  Using dual crystal probe pulse-echo method, echo-echo method (UM-1D)
 Standard Mode Measurement Range  0.8mm~300mm (depending on the probe used and the material being measured)
 Coating Mode Measurement Range  3mm~20mm (using standard probe, UM-1D)
 Measurement Lower Limit for Pipe Material (Steel)  Middle 20mm×3.0mm (standard probe), Middle 15mm×2.0mm (small diameter pipe probe), measurement error ≤±0.1mm.
 Measurement Error  ±0.1mm H<10mm:±(1%H+0.1)mm H≥10mm Note: H is the actual thickness of the measured object
 Repeatability  0.1mm
 Measurement Resolution  0.1mm or 0.001in
 Unit  Millimeters or Inches
 Calibration Mode  Quick Calibration, Two-Point Calibration
 V Path Correction  Automatic V sound path correction, compensating for the non-linearity of the dual crystal probe
 Measurement Update Rate  4Hz
 Material Sound Speed Range  1000~9999 m/s or 0.0394~0.3937 in/us
 Storage Function  Divided into 5 arrays, can store 500 thickness values
 Minimum Value Capture  Refers to capturing the minimum value in the measurement process
 Alarm Function  Can set thickness limits, with alarm sound prompts for measurements outside the limits
 Sound Speed Storage  Can store sound speeds of 5 different materials
 Working Language  Can choose Chinese, English
 Communication Interface  USB interface or RS232 serial port (optional)
 Application Software  UMView software for data transmission, storage, analysis, and document editing (optional)
 Power Supply  Two 1.5V AA batteries, with low voltage prompt when power is insufficient
 Operating Time  Alkaline batteries can be used for 200 hours (without backlight)
 Automatic Shutdown  Automatically shuts down after 5 minutes of inactivity
 Working Temperature  -10℃~+50℃
 Size  149mm×73mm×32mm (H×W×D)
 Weight  210g (including battery)